Say goodbye to handling and processing samples for purge-and-trap analysis of volatile organic compounds (VOCs). The 4100 Sample Processor is designed to operate with (single or dual) Eclipse 4760 purge-and-trap sample concentrators. You can trust the VOA Constrictor™ vial gripper technology to carefully handle samples, while high-speed injection valves improve the precision of standard addition, reduces the volume of standards used, and decreases laboratory operating costs.
Model 4100 Features:
- Holds 100 samples in two 50-position vial racks
- Operates with one or two Eclipse 4760 purge-and-trap instruments
- Vial cooling option keeps vials at 10 °C or less in accordance with USEPA Method 524.3
- VOA Constrictor™ vial gripper has built-in vial sensor
- Standard Addition Module has two 3-mL reservoirs, expendable to four 3-mL reservoirs
- High-speed injection valves in Standard Addition Module inject programmed volume with no excess or waste
- VOA View Windows®-based software offers an intuitive user interface with a real-time view of sampling progress
General Specifications | |
---|---|
Purge-and-Trap Compatibility | Operates with one or two Eclipse 4660 instruments |
Sample Capacity | 100 standard USEPA 40-mL VOA Vials |
Sample Loop | 5mL standard loop, 10 or 25 mL loops optional |
Automated Standard Addition | (2) 3mL reservoirs standard, expansion to (4) optional |
Sampling Mode – Water | Aspiration and transfer to purge-and-trap instrument sparge vessel |
Sampling Mode – Soil | In-vial needle sparging and transfer to trap of purgeand-trap instrument |
Sample Transfer Pathway – Water Samples |
Silcosteel® 316L stainless steel and PEEK® |
Sample Transfer Pathway – Soil Samples |
Silcosteel® 316L stainless steel and PEEK® |
Blank Water Transfer | PFA–Teflon® |
Vial Trays | (2) 50-position VOA vial trays |
Vial Gripper Mechanism | Pneumatically actuated cylindrical gripper with vial sensing |
Vial Cooling Option | Cooling base for two 50-position vial trays with quickdisconnect fittings for recirculating chiller |
X/Y-Axis Mechanism | Digitally controlled linear drive rails |
Z-Axis Mechanism | Pneumatically actuated |
Operating Software | VOA View Windows®-based GUI |
Operating System | Windows® 7or Windows® 8 and 8.1 |
PC to 4100 Communications | USB port for each 4100 instrument connected to PC |
Water Supply | Clean, VOC-free water for rinsing sample pathway |
Gas Requirements – Sample Transfer |
99.999% (UHP Grade) He or N2 |
Gas Requirements – Gripper / Z-axis Actuator | Nitrogen |
Power Requirements | 110VAC / 60HZ or 230VAC / 50Hz |
Dimensions | 27 in. W x 24 in. D x 21.5 in. H 68.6 cm W x 61 cm D x 54.6 cm H |
Weight | 36 Kg (80 lbs) , typical depending on options |
Certifications | CE Safety and EMC EN50082-1/EN55011 Group 1 Class A |